The SC-5, SC-4, SC-3, SC-2, and SC-1 SimCube® NIBP Simulators are compact, affordable and easy to use. These hand-held simulators quickly and accurately determine whether or not a monitor's NIBP feature is working to spec. As cuff pressure changes, SimCube simulates real-life oscillometric pulses by varying both the size and shape of the wave. Get 150-250 NIBP simulations on compatible SimCube models using the optional Battery Boost with a set of 4 "AA" alkaline or NiMH batteries!
Ultra-portable in a 3" X 3" X 3-1/2" (7.6 x 7.6 x 8.9 cm) package, SimCube is constructed of 1/8"(3.2mm)-wall extruded aluminum for superior strength and corrosion resistance. Encased in a vinyl jacket for scratch resistance and quiet operation. Tested for ruggedness — 50 drops from a height of 3.5 feet (1.1 m) to a hard surface with no failures.
SC-5 SimCube® NIBP Simulator with Peak Detect, ECG, Pacer, Respiration, Alarm Test, IBP, Arrhythmia, and Leak Test
The SimCube SC-5 combines the functions of both a manometer (pressure gauge) and an NIBP simulator. The SC-5 NIBP function has five modes: Manometer (+/- 400mmHg), Adult (120/80 mmHg), Neonatal (70/40 mmHg), Hypertensive (190/120 mmHg), and Hypotensive (80/40 mmHg). A single button toggles through each of the modes. A mini-DIN connector provides access to 3 dynamic IBP values (120/80, 70/40, 190/120) and 7 static IBP values (0, 25, 50, 100, 150, 200, 250). Ten snaps provide 3-, 5-, or 12-lead electrocardiogram (ECG) simulation at 3 heartrates (70, 95, Asystole) plus a Heartrate sequence, as well as Arrhythmia, Pacer, and 3 Resp rates with a Resp Sequence. An Alarm Test mode tests heart rate alarms. Peak Detect (Overpressure) measurement and Leak Rate measurement functions are provided. The SC-5 is compatible with theBattery Boost.
SC-4 SimCube® NIBP Simulator with Peak Detect, ECG, Pacer, Respiration, and Alarm Test
The SC-4 NIBP function has the following modes: Manometer (0-400mmHg), Adult (120/80 mmHg), Neonatal (70/40 mmHg), and Hypertensive (190/120 mmHg). A single button toggles through each of the modes. Ten snaps provide 3-, 5-, or 12-lead electrocardiogram (ECG) as well as Pacer, Resp, a Heartrate/Resp sequence and an Alarm Test mode. Peak Detect (Overpressure) measurement function is provided. The SC-4 is compatible with theBattery Boost.
The SC-3 NIBP function has the following modes: Manometer (0-400mmHg), Adult (120/80 mmHg), Neonatal (70/40 mmHg), and Hypertensive (190/120 mmHg). A single button toggles through each of the modes. The Peak Detect (Overpressure) mode measures the peak pressure produced by the patient monitor. The readout latches the highest pressure value sensed, in 0.5mmHg steps. The SC-3 is compatible with theBattery Boost.
SC-2 SimCube® NIBP Simulator with ECG, Pacer and Respiration
Why carry a manometer and separate NIBP and ECG/Resp simulators when you can have them all combined in a convenient, compact package? This configuration of SimCube includes four NIBP modes: Manometer (0-400mmHg), Adult (120/80 mmHg), Neonatal (70/40 mmHg), and Hypertensive (190/120 mmHg), with the addition of 10 snaps for 3-, 5-, or 12-lead electrocardiogram (ECG) simulation. Pacer simulation and respiration simulation allow testing of primary patient monitor hardware circuits.
The basic SimCube combines the functions of both a manometer (pressure gauge) and an NIBP simulator. SimCube's NIBP function has four modes: Manometer (0-400mmHg), Adult (120/80 mmHg), Neonatal (70/40 mmHg), and Hypertensive (190/120 mmHg). A single button toggles through each of the modes.