(click here for a PDF file of all specs)
Technical Specifications for SimCube SC-1 and SC-2
(click here for a PDF file of specs for SC-1 and SC-2)
Physical Dimensions
SC-1/2
Size
3”x 3”x 3.5” (7.6cm X 7.6cm X 8.9cm)
Weight
2.5 Lbs
Power
External A/C Adaptor (Output: 6VDC / 2amps, 2.1mm, center positive connector)
NIBP Connection
Quick Disconnect
ECG/Resp Connection
10 ECG snaps
Manometer
Range
0 to +480 mmHg
Precision
0.5 mmHg
Accuracy
+/-1% of reading
User Interface
Single Button Operation
Operating Modes
Manometer
Adult NIBP
Neo NIBP
Hypertensive NIBP
ECG Asystole (SC-2)
ECG Resp OFF (SC-2)
ECG Pace ON (SC-2)
NIBP Adult Simulation
Simulated Pressure
120/80 (97) mmHg
Simulated Heart Rate
70 bpm
Simulated Pulse Volume
1 ml
NIBP Neonatal Simulation
Simulated Pressure
70/40 (51) mmHg
Simulated Heart Rate
94 bpm
Simulated Pulse Volume
0.25 ml
NIBP Hypertensive Simulation
Simulated Pressure
190/120 (142) mmHg
Simulated Heart Rate
70 bpm
Simulated Pulse Volume
1 ml
ECG Simulation
(SC-2 only)
Isolated
Yes
Synchronized with NIBP
No
R Wave Size
1mV (lead I) +/- 5%
R Wave Width
35 ms
Wave Shape
R wave only
Connection
10 Snaps
Simulation Rates
70 bpm (94 bpm for neonatal), Asystole, Pacer
Pacer Simulation
(SC-2 only)
Isolated
Yes
Synchronized with NIBP
No
Pacer Size
3 mV
Pacer Width
1.2 ms
Respiration Simulation
(SC-2 only)
Isolated
Yes
Wave Shape
Square Wave
Size
4 Ohm
Rate
35 bpm (47 bpm for neonatal)
Environmental
Voltage Range
100-240 VAC, 50-60 Hz
Technical Specifications for SimCube SC-3 and SC-4
(click here for a PDF file of specs for SC-3 and SC-4)
Physical Dimensions
SC-3/4
Size
3”x 3”x 3.5” (7.6cm X 7.6cm X 8.9cm)
Weight
2.5 Lbs
Power
External A/C Adaptor (Output: 6VDC / 2amps, 2.1mm, center positive connector) or 4 AA Batts (with Battery Boost Option)
NIBP Connection
Quick Disconnect
ECG/Resp Connection
10 ECG snaps
Manometer
Range
0 to +480 mmHg
Precision
0.5 mmHg
Accuracy
+/-1% of reading
User Interface
Single Button Operation
Operating Modes
Adult NIBP
Neo NIBP
Hypertensive NIBP
Manometer
Peak Detect / Overpressure
Alarm Test (SC-4)
ECG Asystole (SC-4)
ECG Resp OFF (SC-4)
ECG Pace ON (SC-4)
NIBP Adult Simulation
Simulated Pressure
120/80 (97) mmHg
Simulated Heart Rate
70 bpm
Simulated Pulse Volume
1 ml
NIBP Neonatal Simulation
Simulated Pressure
70/40 (51) mmHg
Simulated Heart Rate
95 bpm
Simulated Pulse Volume
0.25 ml
NIBP Hypertensive Simulation
Simulated Pressure
190/120 (142) mmHg
Simulated Heart Rate
70 bpm
Simulated Pulse Volume
1 ml
ECG Simulation
(SC-4 only)
Isolated
Yes
Synchronized with NIBP
Yes
R Wave Size
1mV (lead I) +/- 5%
R Wave Width
35 ms
Wave Shape
QRS wave
Connection
10 Snaps
Simulation Rates
70 bpm (95 bpm for neonatal), Asystole, Pacer
HR Sequence (Alarm Test)
(SC-4 only)
30 seconds each of: 30, 60, 90, 120, 45, 160, and 220 bpm
Pacer Simulation
(SC-4 only)
Isolated
Yes
Synchronized with NIBP
Yes
Pacer Size
3 mV
Pacer Width
1.2 ms
Respiration Simulation
(SC-4 only)
Isolated
Yes
Wave Shape
Square Wave
Size
4 Ohm
Rate
35 bpm (47 bpm for neonatal)
Simulation rates
30 seconds each of: Apnea, 30, 45, 60, 22, 80, 110.
Environmental
Voltage Range
100-240 VAC, 50-60 Hz
Technical Specifications for SimCube SC-5
(click here for a PDF file of specs for SC-5
)
Physical Dimensions
SC-5
Size
3”x 3” x 3.5” (7.6cm X 7.6cm X 8.9cm)
Weight
2.5 Lbs
Power
External A/C Adaptor (Output: 6VDC / 2amps, 2.1mm, center positive connector) or
4 AA Batteries (with Battery Boost Option)
NIBP Connection
Quick Disconnect
ECG/Resp Connection
10 ECG snaps
IBP Connection
Mini-DIN
Manometer
Range
– 400 to + 400 mmHg
Precision
0.1 mmHg
Accuracy
+/-1 mmHg
User Interface
Single Button Operation
Operating Modes
Adult NIBP
Neo NIBP
Hypertensive NIBP
Hypotensive NIBP
Manometer
Peak Detect / Overpressure
HR Seq. Alarm Test
ECG Asystole
ECG Resp OFF
ECG Pace ON
Arrhythmia Sequence
Invasive BP Zero
Invasive BP 100, 200
Invasive BP Sequence
NIBP Adult Simulation
Simulated Pressure
120/80 (97) mmHg
Simulated Heart Rate
70 bpm
Simulated Pulse Volume
1 ml
NIBP Neonatal Simulation
Simulated Pressure
70/40 (51) mmHg
Simulated Heart Rate
95 bpm
Simulated Pulse Volume
0.25 ml
NIBP Hypertensive Simulation
Simulated Pressure
190/120 (142) mmHg
Simulated Heart Rate
70 bpm
Simulated Pulse Volume
1 ml
NIBP Hypotensive Simulation
Simulated Pressure
80/40 (58) mmHg
Simulated Heart Rate
70 bpm
Simulated Pulse Volume
1 ml
ECG Simulation
Isolated
Yes
Synchronized with NIBP
Yes
R Wave Size
1mV (lead I) +/- 5%
R Wave Width
35 ms
Wave Shape
QRS wave
Connection
10 Snaps
Simulation Rates
70, 95, Asystole, Arrhythmia, Pacer, HR seq.
HR Sequence (Alarm Test)
30 seconds each of:
30, 60, 90, 120, 45, 160, and 220 bpm
Pacer Simulation
Isolated
Yes
Synchronized with NIBP
Yes
Pacer Size
3 mV
Pacer Width
1.2 ms
Respiration Simulation
Isolated
Yes
Wave Shape
Square Wave
Size
4 Ohm
Rate
20 bpm (40 bpm for neonatal)
Simulation rates
30 seconds each of: Apnea, 30, 45, 60, 22, 80, 110.
Arrhythmia Simulation
Cardiac failure sequence: 90 seconds of normal beats interspersed with PVCs and Runs, followed by 30 seconds of VTAC, followed by 30 seconds of VFIB, concluding with 30 seconds of asystole.
Peak Detect
Precision
1mmHg
Invasive Blood Pressure Simulation
Isolated
Yes
Synchronized with NIBP
Yes
Excitation Voltage
DC range = 3.3 to 5.7 AC range = 6.65 to 11.4p-p
Pressure range
0-250 mmHg
Simulated Pressure accuracy
+/- 1 mmHg
Wiring
+ Excit = pin 1, - Excit = pin 4, + Sig = pin 3, -Sig = pin 6
Simulation rates
Dynamic = 120/80, 70/40, 190/120.
Static = 0, 100, 200.
Step = 0,25,50,100,150,200, 250
Environmental
Voltage Range
100-240 VAC, 50-60 Hz