Features:
Accuracy |
1% of reading ± 1 LSD for 1ml/hr to 999 ml/hr, 2% ± 1 LSD below 1ml/hr. 0.85% of reading for 500ml/hr, 12ml Volume under steady flow conditions. |
Range |
-28° to 115.5° C |
Resolution |
Flow Rates: 0.01ml; Volume: 0.001ml |
Occlusion Pressure Testing |
Provides Real Time and Peak Pressure Measurements |
Digital Timer/Stopwatch |
Automatically starts and stops with fluid measurements. No need to stop analyzer when test complete. |
Temperature Meter Features:
Range |
-28.9° to 115.5° C |
Units of Measure |
Fahrenheit, Celsius |
General Features:
Power |
6VDC DC Adapter or (2) AA Batteries |
Battery Life |
Alkaline: Min 4 hours Lithium: Min. 12 hours |
Warranty |
4-Year Limited Warranty |
Included Accessories |
60cc Syringe, 3-way Stopcock, Temperature Probe Extension Cable, Pressure Testing Luer Connector, Cleaning Solution, Flush Bottle, Cleaning Brush, Output hose, USB Cable, Carrying Case, DC Power Supply |
Optional Accessories |
IV Pole Mount |
Flow and Volume Performance:
Range |
0.5 to 999 ml/hr |
Accuracy |
1% of reading ± 1 LSD for 1ml/hr to 999ml/hr, 2% ± 1 LSD below 1ml/hr. 0.85% of reading for 500ml/hr, 12 ml Volume under steady flow conditions. |
Requirement for Accuracy:
Flow Rate |
Required Volume/Time |
.5 to 10ml/hr |
0.5ml |
10.1 to 199ml/hr |
3 minutes (see 3 Minute Testing Quick Reference chart) |
200 to 999ml/hr |
10ml |
Precision:
Precision of FlowTrax is equivalent to 0.01 g based on the conversion of volume (ml) to weight (g) by calculation using standard temperature and density value of water. Precision was established using a rate
of 500ml/hr and a “dispensed” volume of 10ml.
Pressure Performance:
Range |
-12.5 to 75 PSI |
Accuracy |
0.5% of reading ± 1 LSD |
Resolution |
0.01 PSI |
Units of Measure |
mmHg, PSI (gauge), inH2O, cmH2O, kPa, atm, inHg, BAR (absolute), BAR (gauge) |
Temperature Performance:
Range |
-28.9° to +115.5° C |
Accuracy |
0.1% ± 0.2 degrees C |
Resolution |
0.1° C |
Digital Timer/Stopwatch:
Range |
0 to 12 hours |
Accuracy |
0.5% |
General:
Power |
Powered by two AA (Alkaline or Lithium) cells or 6 VDC Input |
Battery Life |
Alkaline: Min. 4 hours Lithium: Min. 12 hours |
“Industry Insider: Pronk Technologies Inc” appeared in 24/7 introducing the FlowTrax IV Pump Analyzer
Clearly demonstrating why someone should pay for a product is not always easy for marketing professionals. The product might function well, but its attributes might be unfamiliar. The product’s attributes might be fine, but they might not be that different from competitors’ products. Greg Alkire, vice president of sales and marketing for Pronk Technologies Inc, Sun Valley Calif, does not come off as if he faces those quandaries often…
[READ MORE]
“The Roundtable: Infusion Therapy Equipment” appeared in TechNation Magazine
This month, TechNation asked infusion therapy equipment experts to weigh in on the latest trends in purchasing and service. Responders include Greg Alkire, Vice President of Sales and Marketing for Pronk Technologies; Andrew Davis, Account Manager for Tenacore Holdings and Dr. Tim Vanderveen, Vice President of the CareFusion Center for Safety and Clinical Excellence….
[READ MORE]
“Pronk Technologies Unveils its Four-in-One Test Device” appeared in 24/7
Pronk Technologies has launched FlowTrax, a palm-sized, multifunction infusion pump analyzer.
According to Pronk, significant technological advances reduce the testing time for servicing and maintaining infusion devices. The new FlowTrax tests infusion pumps for flow rate and volume performance in three minutes or less with a 1% accuracy using FlowTrax’s patent-pending flow tracking technology…[READ MORE]